Lab 7:  Introduction to the Oscilloscope

INTRODUCTION:

          The oscilloscope, or scope, is one of the most widely used electronic instruments.  It can be used to observe and measure voltage signals, either static or time- varying, over a very wide range of amplitudes and frequencies.  The purpose of the present experiment is to give you an introduction to the oscilloscope and several typical applications. 


The Oscilloscope

          An oscilloscope makes a plot of voltage versus time and shows it to you on the screen.  It periodically updates the display, so if the signal changes, the display will show the change.  The screen itself looks a little like a graph.  The horizontal axis is the time and the vertical axis is the voltage, see fig. 1 below. 

The oscilloscope will allow two different signals (voltage sources) to be plotted at the same time, one for each channel.  The oscilloscopes you will be using will also display the scale for the plots, i.e. how much each major division represents.  The scale is at the bottom of the display.  In the figure 2 below, channel 1 has a scale of 5V per major division, and channel 2’s is also 5V per major division.  If the scale is 5V, then a signal 2 ˝ major divisions high represents 2.5´5V = 12.5V.  (Note that the triangular waveform in the figure has a range of about 2.4 divisions from the lowest part to the top, or 2.4 divisions peak to peak.  This means that the peak-to-peak voltage is about 12V.)  Similarly a time scale of 1ms means a trace that extends 3 major divisions horizontally is 3´1ms = 3ms long.  (In the figure above, the scale is 500ms per major division and the length of a cycle is about two divisions, so the period is 1,000ms or 1ms.)  Note that the voltage scales for the two different channels may be different, but the time scale is the same for both.

          There are three sets of controls you can vary.  (See fig. 3.)  They are labeled VERTICAL, HORIZONTAL and TRIGGER on the oscilloscope. 

          The VERTICAL controls are under the vertical heading and allow you to change the voltage scale for each channel by turning the knob labeled VOLTS/DIV.  You can also move the display for the channel up and down on the screen by turning the knob labeled POSITION. 

          The HORIZONTAL controls are under the horizontal heading and allow you to change the time scale by turning the SEC/DIV knob.  You can also offset the display from left to right by turning the POSITION knob.

          A final control (for this lab) is the AutoSet button at the upper right of the scope.  When you press that the oscilloscope tries to find a suitable setting for the voltage and time scales.  It is a handy feature of these scopes.

 

PART I.   DC VOLTAGE MEASUREMENTS

The DC power supplies have two terminals, a positive (red) and a common (Black).   In this experiment, the positive DC voltage will be measured with the oscilloscope.

          Connect a probe to the scope amplifier on the left.  Plug the BNC end of the probe into CH 1 (channel 1) input on the oscilloscope and connect the other end of the probe into the + terminal of the power supply using jumper wires provided.  Connect the ground strap of the probe to the common terminal in the power supply.  Push the POWER  knob to turn on the scope.  Press the AutoSet button and look at the display.  The AutoSet button is in the upper right corner of the scope.  The auto set feature will adjust the settings on the display in an attempt to make the display appropriate to the input. 

Now measure the voltage put out by the power supply for different knob settings on the power supply by noting the vertical deflection on the scope and the vertical sensitivity as you change the voltage.  Try about 4 or 5 different voltages from about 1V to 15V.  Simultaneously make the same measurement on a digital voltmeter.  (You may have to change the volts/div setting on the scope as you change your voltage.  See fig. 3 above.  The VOLTS/DIV knob is just above the place you connect the probe for that channel.)  Make a table of your data showing the voltmeter reading and the oscilloscope reading.

Voltage DMM

Scope Voltage Scale

Divisions

Volts (Scope)

Volts (Scope – Measure)

% diff btw DMM & Scope

 

 

 

 

 

 

 

PART II. AC MEASUREMENTS

THE FUNCTION GENERATOR

A drawing of the function generator is shown below.  It will generate various types of signals at different frequencies.  Two typical signals are sine waves and square waves.  It also allows you to choose the amplitude and frequency of the signals.


 The frequency is determined by the push buttons across the top (left and center ones labeled 1M, 100k, 10k, 1k, 100, 10, and 1).  These select the coarse range and the dial gives finer control of the frequency.  If the dial is at 0.8 and the 1k button is pressed, the frequency is 0.8 x 1000 = 800 Hz.  (k is shorthand for 1000.)

          A drawing of the function generator is shown above.  It will generate various types of signals at different frequencies.  Two typical signals are sine waves and square waves.  It also allows you to choose the amplitude and frequency of the signals.  The frequency is determined by the push buttons across the top (left and center ones labeled 1M, 100k, 10k, 1k, 100, 10, and 1).  These select the coarse range and the dial gives finer control of the frequency.  If the dial is at 0.8 and the 1k button is pressed, the frequency is approximately 0.8´1000 = 800 Hz.  (k is shorthand for 1000.)

          The type of waveform is selected by pushing one of the three buttons in the upper right above the label function.  (But not the last one on the right labeled attenuator!)  They are square wave, triangle wave and sine wave.  The symbol above each mimics one cycle of the waveform. 

          The amplitude is controlled by the dial labeled amplitude.  If it is set at the min. position, the amplitude is not zero, but about 300 mV peak to peak.

A.      Sine Wave.  In this section, a time-varying signal form a signal generator will be investigated.  Connect the BNC cable from the generator output marked MAIN to the probe from CH1 on the oscilloscope.  Push the sine wave button to select a sine wave output.  (It is just to the left of the -20dB button.)  Set the range push button to 1k and turn the frequency dial to 1.0.    Adjust the amplitude dial on the function generator to so it is about 1/4 turn from the minimum.  Press the AutoSet button and you should now observe a sine wave on the screen.

1.       Record your voltage scale.

2.       Determine the peak to peak amplitude and calculate the peak amplitude.

3.       Calculate the RMS amplitude from the peak amplitude.

4.       Use the Measure button on the scope to have it find the peak and RMS.  How does your RMS compare the to scope’s value?

Voltage scale

Div

Pk to Pk Voltage

Peak Voltage.

RMS

From Pk

RMS Scope

 

 

 

 

 

 

 

B.      Square Wave.  In this section, a time-varying signal form a signal generator will be investigated.  Connect the BNC cable from the generator output marked MAIN to the probe from CH1 on the oscilloscope.  Push the square wave button to select a square wave output.  (It is just to the right of the INVERT button.)  Set the range push button to 1k and turn the frequency dial to 1.0.    Adjust the amplitude dial on the function generator to so it is about 1/4 turn from the minimum.  Press the AutoSet button and you should now observe a square wave on the screen.

1.       Record your time scale and measure the period (time for one complete waveform) in horizontal divisions and in time.  Calculate the corresponding frequency. 

2.       Repeat for frequency setting of 100kHz.  (You can change the frequency just by pushing the other frequency range buttons, i.e. the 100k.)  You will need to push the AutoSet button after each change of frequency, or change the time scale manually by turning the knob under HORIZONTAL that is labeled SEC/DIV.

Make a table for these like the one below.  (Remember f = 1/T.)  Watch your units!

Generator frequency (Hz)

Scope div.

Time per div.

Period

Frequency (Hz)

Scope Measurement Freq.

 

 

 

 

 

 

 

 

3.       Timing Measurements

          The Oscilloscope is often used for timing measurements.  For instance if you want to find out how long it takes for an electrical signal to travel from one end of a cable to the other end of the cable, you can use the oscilloscope to measure that.  You can hook up a generator to the cable with a BNC “Tee” with one end of the Tee going to channel 1 on the scope and the other end to the first end of the cable.  The other end can go to channel 2 on the scope.  Let the generator put out a Square Wave of about 200kHz and measure the time difference between when the leading edge of the square wave arrives at channel 1 and when it arrives at channel 2. 

          You can do a similar measurement just using channel 1 if you don’t connect the other end of the cable to anything.  When the wave reached the unattached end, it with reflect back.  You can measure the time between the arrival of the first edge and the arrival of the reflection.  Now the signal has traveled twice the length of the cable (down and back). 

          If you measure the length of the cable, you can calculate the speed of the electromagnetic wave going down the cable. 

          I want you to measure the time it takes for the wave to go down the cable and measure the length of the cable and calculate the wave speed.  You should sketch what you see on the screen of the oscilloscope and label the signals showing where you took the time difference.

          If the wave speed is given by

         

find the value of k for your cable.